mode. Its duration time is 9.5 msec. and its interval is 2 sec.
3. Set channel #2 to the trigger mode.
Press keys: ‘TRIG, 2, ENTER’ (see Section 7)
Nothing happens, this is because Channel #2 is now in the TRIGGER mode, but is waiting to accept
a trigger. You can trigger it in 3 ways:
a) MANUAL TRIGGERING -
Press key: ‘2’.
(‘2’ is the channel number.) Note that each press on ‘2’ delivers the triggered pulse. Press ‘2’ several
times.
b) EXTERNAL TRIGGERING -
Whenever there is an input pulse in the ‘EXT 2’ input it triggers channel #2 (see Section 9).
c) INTERNAL TRIGGERING -
You can internally connect each channel to any number of the other channels (see Section 9).
4. Connect channel #1 internally to channel #2 -
Press keys: ‘CONNECT,1,2,ENTER’ to connected. (see
Section 9).
Each output of channel #1 triggers channel #2.
Notes:
One. You don’t have to worry about the amplitude or polarity of the trigger
source.
Two. The instruction itself is very straightforward (i.e. similar to the way
you would say it: - “CONNECT from #1 to #2, ENTER”).
Three. You can add manual pulses by pressing ‘2’.
5. Set the duration of channel #2 to be 15 msec. -
Press key: DURA,2,15,ENTER,3,ENTER’ (see Section 8)
1 - Introduction