A.M.P.I. Master-8 Instrukcja Użytkownika Strona 6

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 14
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 5
7. Modes of Operation
Each of the 8 channels can operate in one of the following modes:
Mode Description
FREE RUN The channel delivers pulses continuously according to the programmed
duration and interval times. The channel is independent of the other
channels.
TRAIN Following a trigger, the channel delivers a train of pulses according to
the programmed duration and interval times. The number of pulses per
train is set by ‘M’.
TRIG Following a trigger, the channel delivers a single pulse according to the
programmed delay and duration times.
DC The channel is time independent, you can manually turn it on and off.
GATED Channels 1 and 2 can be externally gated via the EXT1 or EXT2 inputs.
The channel delivers pulses continuously while it is externally gated.
When the external input stops, the channel stops working. A new input
reactivates the channel.
In either TRAIN or TRIG mode you can trigger the channel in the following ways:
1. Manually
2. The internal connections
3. EXT1 (for channel 1 only) EXT2 (for channel 2 only)
7.1 Setting the modes of Operation
To set the mode of operation of a channel, first press the selected mode key and then the channel number.
Example 1 : In order to set channel #2 to the FREE-RUN mode -
Press keys : ‘FREE,2,ENTER’.
Channel #2 now delivers pulses continuously according to its duration and interval times (or the default
times).
Example 2: In order to set channel #3 to the TRAIN mode -
Press keys: ‘TRAIN,3,ENTER’ .
Channel #3 is now in the TRAIN mode. You can trigger it in several ways. Trigger it now manually by
pressing ‘3’ (channel number)